Divisions

Oncology > POLYSITE® implantable catheter ports >


POLYSITE® Standard, Titanium 4000 series


Intravenous implantable catheter port
Anatomical shape
Low weight
Easy to connect
MRI-compatible

CHARACTERISTICS OF THE PORT
Internal volume: 0.55 ml
References Type of port/Material material/
Matériaux
With
introducer

ducteur
Without introducer Pre-connected Not pre-connected Base diameter
(mm)
Height
(mm)
Septum diameter
(mm)
(mm)
Weight
(g)
Port outlet diameter
(mm)
4017
4017 ISP
4117
4117 ISP
POM*/Titanium
Adult
Standard
Port
-
X
-
X
X
-
X
-
-
-
X
X
X
X
-
-
35,7
x
22,5
12,1 12,1 7,7 1,0
4008
4008 ISP
4108
4108 ISP
POM*/Titane -
X
-
X
X
-
X
-
-
-
X
X
X
X
-
-
35,7
x
22,5
12,1 12,1 7,7 1,2
* Polyoxymethylene

CHARACTERISTICS OF THE SILICON CATHETERS
5 cm graduations over 40 cm
Radio-opaque
References ORX Catheters
External diameter (mm) Internal diameter
(mm)
Wall thickness
(mm)
Length
(mm)
Internal diameter
ml/10cm
Sheath introducer size Flow rate
(ml/min)*
22G
Flow rate
(ml/min)*
20G
Flow rate
(ml/min)*
19G
4008
4008 ISP
Silicone 2.4 1.2 0.6 600 0.125 8F 13 30 37
4108
4108 ISP
Silicone 2.4 1.2 0.6 600 0.125 8F 13
30 37
*Maximum flow rate in ml/min, with a straight Huber needle of a given size, under the pressure of a one-metre column of water

CHARACTERISTICS OF THE POLYURETHANE CATHETERS
5 cm graduations over 40 cm
Radio-opaque
References ORX
Catheters
External diameter (mm) Internal diameter
(mm)
Wall thickness
(mm)
Length
(mm)
Internal diameter
ml/10cm
Sheath introducer size Flow rate
(ml/min)*
22G
Flow rate
(ml/min)*
20G
Flow rate
(ml/min)*
19G
4017
4017 ISP
Polyurethane 2.1 1.4 0.7 600 0.176 7F 13 31 39
4117
4117 ISP
Polyurethane 2.1 1.4 0.7 600 0.176 7F 13 31 39
*Maximum flow rate in ml/min, with a straight Huber needle of a given size, under the pressure of a one-metre column of water

Commercial brochure - 188 Ko (PDF)