Divisions

Oncology > POLYSITE® implantable catheter ports >


POLYSITE® POM resin 1000 series


Implantable catheter port
Flexible and reinforced outlet tube, anti-bending system, optimal safety.
Radio-opaque, self-sealing silicone septum allowing more than 1,000        punctures per cm²
Simple and safe port/catheter connector
Non-thrombogenic, radio-opaque, silicone or polyurethane catheter

CHARACTERISTICS OF THE PORTS: VENOUS
Reference Type of port With
peel-off introducer
With non-peel-off introducer Without introducer Base diameter
(mm)
Port diameter
(mm)
Port height
(mm)
Septum diameter
(mm)
Septum thickness
(mm)
Internal volume
(ml)
Weight
(g)
1007
1007 IS
1007 ISP
Microport
Adult
-
-
7J
-
7J
-
X
-
-
31x22,8 17,6 10,7 10,3 4,8 0,25 4,5
1004 Paediatric - - X 31x22,8 17,6 10,7 10,3 4,8 0,25 4,5
1031
1031 IS
1031 ISP
High-flow rate Micro-port
Adult
-
-
10J
-
10J
-
X
-
-
31x22,8 17,6 10,7 10,3 4,8 0,3 4,8
1040
1040 ISP
Pre-connected
Micro-port
Adult
-
7J
-
-
X
-
31x22,8 17,6 10,7 10,3 4,8 0,25 4,5
The POLYSITE implantable port is double packed

CHARACTERISTICS OF THE PORTS: SPINAL
Reference Type of port With
peel-off introducer
With non-peel-off introducer Without introducer Base diameter
(mm)
Port diameter
(mm)
Port height
(mm)

(mm)
Septum diameter
(mm)
Septum thickness
(mm)
Internal volume
(ml)
Weight
(g)
1017 Micro-port - - X 31x22,8 17,6 10,7 10,3 4,8 0,25 4,5
The POLYSITE® implantable port is double packed.

CHARACTERISTICS OF THE PORTS: ARTERIAL
Reference Type of port With
peel-off introducer
With non-peel-off introducer Without introducer Base diameter
(mm)
Port diameter
(mm)
Port height
(mm)
Septum diameter
(mm)
Septum thickness
(mm)
Internal volume
(ml)
Weight
(g)
1019 Micro-port - - X 31x22,8 17,6 10,7 10,3 4,8 0,25 4,5
The POLYSITE® implantable port is double packed.

CHARACTERISTICS OF THE PORTS: PERITONEAL
Reference Type of port With
peel-off introducer
With non-peel-off introducer Without introducer Base diameter
(mm)
Port diameter
(mm)
Port height
(mm)
Septum diameter
(mm)
Septum thickness
(mm)
Internal volume
(ml)
Weight
(g)
1016 Standard
Adult
- - X Diam.30 22 16,2 11,5 6,2 0,8 10,8
The POLYSITE® implantable port is double packed.

CHARACTERISTICS OF THE SILICONE CATHETERS: INTRAVENOUS
1 cm graduations with numbered markings every 5 cm from 5 to 30 cm
Radio-opaque
  References ORX catheter External diameter (mm) Internal diameter
(mm)
Wall thickness
(mm)
Length
(mm)
Internal volume
ml/10 cm
Flow rate
(ml/min)*
Micro-port Adult 1007
1007 IS
1007 ISP
Silicone 2,16 1,02 0,57 600 0,1 12 (22g)
23 (20g)
26 (19g)
Paediatric 1004 Silicone 1,65 0,8 0,425 600 0,06  
High-flow rate Micro-port
Adult
1031
1031 IS
1031 ISP
Silicone 3,18 1,57 0,805 600 0,22 15 (22g)
29 (19g)
Pre-connected
Micro-port
Adult
1040
1040 ISP
Silicone 2,16 1,02 0,57 600 0,1  
*Maximum flow rate in ml/min, under the pressure of a one-metre column of water

CHARACTERISTICS OF THE POLYURETHANE CATHETERS: INTRASPINAL
Sealed extremity
Graduations over 15 cm
Side holes
Mandril
Radio-opaque
  References ORX catheter External diameter (mm) Internal diameter
(mm)
Wall thickness(mm)
Length
(mm)
Internal volume
ml/10 cm
Flow rate
(ml/min)*
Micro-port 1017 Polyurethane 0,8 0,5 0,15 800 0,028  
*Maximum flow rate in ml/min, under the pressure of a one-metre column of water

CHARACTERISTICS OF THE SILICONE CATHETERS: INTRAARTERIAL
Anti-reflux valve
Radio-opaque
  References ORX catheter External diameter (mm) Internal diameter
(mm)
Wall thickness
(mm)
Lenght
(mm)
Internal volume
ml/10 cm
Flow rate
(ml/min)*
Standard
Adult
1019 Silicone 2,3 0,8 0,75 600 0,06  
*Maximum flow rate in ml/min, under the pressure of a one-metre column of water

CHARACTERISTICS OF THE SILICONE CATHETERS: INTRAPERITONEAL
Multi-perforated
Radio opaque
  References ORX catheter External diameter (mm) Internal diameter
(mm)
Wall thickness
(mm)
Length
(mm)
Internal volume
ml/10 cm
Flow rate
(ml/min)*
Standard
Adult
1016 Silicone 4,8 2,8 1,0 500 0,66  
*Maximum flow rate in ml/min, under the pressure of a one-metre column of water